247-L
247-L
End of lifeConductivity and Temperature Probe
No longer available to order
This product is retained for service, documentation, and compatibility information.
Replaced by
- CS547A
Water Conductivity and Temperature Probe
- Model:
- 247-L
- Item Number:
- 247-L
Overview
The 247 was similar to the 247W-L, except it did not have a stainless steel sleeve, and therefore had a smaller diameter. The 247W measured conductivity using three passivated 316 stainless steel annular electrodes with dc isolation capacitors. A thermistor measured temperature.
Technical Description
The 247 had a pH range of 3.0 to 9.0, with a conductivity accuracy of ±5% (at 25°C) over 0.45 to 7.0 mS/cm range, ±10% (at 25°C) over 0.005 to 0.45 mS/cm range. Temperature polynomial linearization accuracy was typically less than ±0.1°C over 0° to +48°C range. Interchangeability accuracy was typically less than ±0.2°C over 0° to +50°C range.
The 247-L (pn 7343) was originally replaced with the CS547, which was replaced with the CS547A.
Specifications
- Operating Temperature
- 0° to 50°C
- Diameter
- 1.91 cm
(0.75 in.)
- Length
- 7.938 cm
(3.125 in.)
- Weight
- 0.2 kg (0.5 lb) with 3.05-m (10-ft) cable
Other Accessories
(2)10060
Calibration Solution for CS547A, 247, or 247W
Calibration solution (NIST-traceable or non-certified) is available for existing 247 customers.
Orderable7421
Split Mesh Cable Grip, 0.230 to 0.320
The 7421 Split Mesh Cable grip is recommended for use in wells to center the cable and to provide a method of suspending the cable, reducing cable stretch.
Orderable
Similar Products
(1)Compatibility
The 247 required one differential analog input, one single-ended analog input, and two excitation channels.
Related Resources
Case Studies
Documents
Services Available
- Repair
- No
- Calibration
- No
- Free Support
- No