39022
Recommended Preventive Maintenance for WEATHERPAK M
No image available
39022
OrderableRecommended Preventive Maintenance for WEATHERPAK M
Overview
The 39022 provides recommended annual preventive maintenance service for the WEATHERPAK™M to ensure these systems are performing within specified accuracies.
Technical Description
This preventive maintenance service provides the following:
- Replacement of all O-rings
- Replacement of barometer hard hat filter
- Temperature/relative humidity service to check the accuracy in room conditions if it has been more than two years since the last service:
- For the HMP50 or HMP60, a replacement relative humidity chip and filter cap are provided.
- For the HC2A-S3 or similar, three-point calibration is performed. The filter cap is inspected and replaced as needed.
- If the sensor cannot be calibrated, an estimate for replacement will be provided.
- Replacement of the WEATHERPAK M lithium battery or batteries
- Replacement of the desiccant packs
- Full testing of the WEATHERPAK M:
- Radio output power test
- A check that the barometric pressure is within spec
- GPS functionality
- Compass test/alignment
- Basic tower test if applicable
- Full testing of the WindSonic/MetSENS200 wind sensor
- Display:
- Verification of communications with the WEATHERPAK M station
- Replacement of the internal batteries as needed