CS215-L
CS215-L
RetiredDigital Air Temperature and Relative Humidity Sensor
No longer available to order
This product is retained for service, documentation, and compatibility information.
Replaced by
- HygroVue 5
Digital Temperature and Relative Humidity Sensor
- Model:
- CS215-L
- Item Number:
- 18007
- Warranty:
- 1 year
Overview
The CS215 is a temperature and relative humidity probe that uses a Swiss-made digital humidity and temperature element that makes accurate and stable measurements. The element is based on Sensirion's CMOSens technology, which has been tested for more than two years in alpine conditions. The CS215 outputs an SDI-12 signal that is measurable by most Campbell Scientific data loggers.
Key Advantages
General-purpose temperature and RH probe
Technical Description
The CS215 uses the Sensirion SHT75, a combined relative humidity and temperature element, to provide accurate, stable measurements. The Sensirion SHT75 element is field-replaceable, eliminating the downtime typically required for the recalibration process. The CS215 outputs an SDI-12 signal that's measurable by many Campbell Scientific data loggers.
The CS215 should be housed in a solar radiation shield—typically the 41303-5A. The 41303-5A 6-plate naturally aspirated shield attaches to a mast, crossarm, or tower leg.
Product Notes
Benefits
- Accurate and stable measurements
- Field changeable element allows on-site recalibration
- Each sensor element is individually calibrated so no further adjustment of the probe is required
- Low power consumption
- Digital SDI-12 output
- Wide operating range
Specifications
- Sensing Element
- Sensirion SHT75
- Measurement Description
- Temperature, relative humidity
- Signal Type/Output
- SDI-12
- Communication Standard
- SDI-12 V1.3
(responds to a subset of commands)
- Housing Material
- Anodized aluminum
- Housing Classification
- IP65
(NEMA 4)
- Sensor Protection
- Outer glass-filled polypropylene cap. Inner expanded PTFE filter.
Filter material has a porosity of 64% and a pore size of < 3μm.
- Supply Voltage
- The supply voltage is typically powered by the data logger's 12 V supply.7 to 28 Vdc
(for serial numbers E13405 and newer)
6 to 18 Vdc(for older models)
- Typical Current Drain
- 120 μA
(quiescent)
1.7 mA(measurement takes 0.7 s)
- EMC Compliance
- Tested and conforms to IEC61326:2002.
- Operating Temperature Range
- -40° to +70°C
- Field Replaceable Chip or Recalibrate
- Field-replaceable chip
- Diameter
- 1.2 cm
(0.5 in.) at sensor tip
1.8 cm(0.7 in.) at cable end
- Length
- 18.0 cm
(7.1 in.) including strain relief
- Weight
- 150 g
(5.3 oz) with 3.05 m (10 ft) cable
Relative Humidity
- Measurement Range
- 0 to 100% RH
(-20° to +60°C)
- Output Resolution
- 0.03% RH
- Accuracy
- ±2%
(10% to 90% range) at 25°C
±4%(0% to 100% range) at 25°C
- Short-Term Hysteresis
- < 1% RH
- Temperature Dependence
- Better than ±2%
(-20° to +60°C)
- Typical Stability
- ±1.0% per year
- Response Time with Filter
- < 20 s
(63% response time in still air)
- Calibration Traceability
- NIST and NPL standards
Air Temperature
- Measurement Range
- -40° to +70°C
- Output Resolution
- 0.01°C
- Accuracy
- ±0.3°C
(at 25°C)
±0.4°C(5° to 40°C)
±0.9°C(-40° to +70°C)
- Response Time with Filter
- < 120 s
(63% response time in air moving at 1 m s-1)
Common Accessories
(3)RAD06
6-Plate Solar Radiation Shield
When exposed to sunlight, the CS215 must be housed in a radiation shield. The RAD06 uses a double-louvered design that offers improved sensor protection from driving rain, snow, and insect intrusion. It also has lower self-heating in bright sunlight combined with higher temperatures (> 24°C (~75°F)) and low wind speeds (< 2m/s (~4.5 mph)), providing a better measurement.
Orderable41303-5A
6-Plate Solar Radiation Shield
When exposed to sunlight, the CS215 must be housed in a radiation shield. The 41303-5A 6-Plate Gill radiation shield can be attached to a mast, tower leg, or CM202-series crossarm.
Please note that if the application requires that the entire body of the CS215-L sensor be engulfed in the radiation shield, the 41003-5 10-plate radiation shield can be used. Use of the 41003-5 also requires purchase of the 6637 adapter.Orderable39642
SDI-12 DIN Rail Terminal Bus Kit without Mounting Bracket
The 39642 makes it easier for you to connect wires from multiple SDI-12 sensors into one set of terminals on a data logger. The 39642 uses six DIN-rail mountable (pn 27479), 6-wire terminal blocks. The 39642 is a jumper that electrically connects two terminal blocks together at a time (pn 32918). This setup allows up to 11 SDI-12 sensors to be connected to one set of terminals on a data logger.
Orderable
Compatibility
Sensor Mounts
When exposed to sunlight, the CS215 must be housed in a 41303-5A radiation shield. To attach the 41303-5A to a CM202, CM204, or CM206 crossarm, place the 41303-5A’s U-bolt in the bottom holes. To attach the radiation shield directly to a tripod mast, tower mast, or tower leg, place the U-bolt in the side holes.
The CS215-L Temperature and RH sensor can also be housed in the 43502 if a special adapter from R. M. Young (R. M. Young pn 43537) is used.
Related Resources
Case Studies

France: Dynamic Agrivoltaism
Measurement of water balance variables with agrivoltaic devices
France (Montpellier – Tresserre – Mallemort – Piolenc)
December 17, 2019
View
Réunion Island: Developing an Ecocity for 140,000 Inhabitants Using Weather Station Data
Collecting and monitoring climatic data for a sunstainable tropical island ecocity
Réunion Island
January 18, 2022
View
Australia: Tropical Cyclone Debbie
Portable anemometer network for wind speed and measurements of land-falling tropical cyclones, to better characterise extreme wind events and their impact on buildings.
North Queensland Coast
July 24, 2017
View
Documents
- Brochure
CS215 Temperature and Relative Humidity Probe
March 2026
View Document - Manual
SDI-12 Terminal Bus Kit Getting Started Guide
May 2026
View Document - Technical Paper
SDI-12 Sensor Troubleshooting
November 2023
View Document - Manual
SDI-12 Terminal Bus Kit Getting Started Guide (online)
May 2026
View Document
Showing 4 of 5 results
See all DocumentsVideos and Tutorials
Services Available
- Repair
- No
- Calibration
- No
- Free Support