3WHB10K
3WHB10K
Orderable10-kohm, Three-Wire Half-Bridge Terminal Input Module
Overview
The 3WHB10K provides the completion resistor required for a three-wire half-bridge measurement. It is typically used with 100 Ω and 1000 Ω platinum resistive thermometers (PRTs).
Technical Description
3WHB10K Schematic

3WHB10K-to-Sensor Hookup

Benefits
- Makes accurate measurement of 3-wire half bridges simple
- Easy to install or remove
- Connects directly into data logger input terminals
- Compatible with most Campbell Scientific data loggers
Specifications
- Used With
- 100 Ω or 1 kΩ platinum resistive thermometer (PRT) or other 3-wire half bridge
- Resistance
- 10 kΩ
(bulk metal foil)
- Tolerance
- ±0.01%
(@ 25°C)
- Maximum Temperature Coefficient
- ±0.8 ppm/°C
- Power Rating
- 0.25 W
(@ 70°C)
- Dimensions
- 1.5 x 1.5 x 2.7 cm
(0.6 x 0.6 x 1.0 in.) for body with prongs
- Weight
- 6 g
(0.2 oz)
Similar Products
(10)
CURS100
100-Ohm Current Terminal Input Module
Orderable
VDIV10:1
10-to-1 Voltage Divider Terminal Input Module
Orderable
VDIV2:1
2-to-1 Voltage Divider Terminal Input Module
Orderable
4WFBS120
120-Ohm, Four-Wire Full-Bridge Terminal Input Module
Orderable
4WFBS350
350-Ohm, Four-Wire, Full-Bridge Terminal Input Module
Orderable
4WFBS1K
1-kohm, Four-Wire, Full-Bridge Terminal Input Module
Orderable
4WPB100
100-Ohm, Four-Wire PRT Bridge Terminal Input Module
Orderable
4WPB1K
1-kohm, Four-Wire PRT Bridge Terminal Input Module
Orderable
4WHB10K
10-kohm, Four-Wire Half-Bridge Terminal Input Module
Orderable
4WPB500
500-Ohm, Four-Wire PRT Bridge Terminal Input Module
Orderable
Compatibility
Note: The following shows notable compatibility information. It is not a comprehensive list of all compatible or incompatible products.
Multiplexers
The AM16/32-series and AM25T multiplexers are compatible with the 3WHB10K.
Data Logger Considerations
The 3WHB10K uses two adjacent single-ended analog inputs; an adjacent analog ground channel accepts the "ground prong" of the 3WHB10K. Not all data logger terminal strips have this sequence on all channels, so please check your data logger's wiring panel to confirm channel assignments, especially if you plan to use multiple terminal input modules.
Newer CR10X Wiring Panels can accommodate both the 3WHB10K and the cover associated with the CR10XTCR (reference thermistor for thermocouple measurements). Specifically, the physical configuration of the thermocouple cover, the data logger input terminals, and the terminal input modules allow sufficient room for the cover to be used with a Terminal Input Module attached. In contrast, older CR10TCRs are not deep enough.
Compatible Products
(16)
Basic Datalogger
Basic Datalogger
RetiredNotes—
Basic Datalogger
Basic Datalogger
RetiredNotes—
CR6
High-Capacity Data Acquisition System
OrderableNotes—
CR10
Measurement and Control System
RetiredNotes—
CR300
Measurement and Control Datalogger
RetiredNotes—
CR310
Measurement and Control Datalogger
OrderableNotes—
CR350
Measurement and Control Datalogger
OrderableNotes—
CR800
Measurement and Control Datalogger
RetiredNotes—
CR850
Measurement and Control Datalogger
RetiredNotes—
CR1000
Measurement and Control Datalogger
RetiredNotes—
CR3000
Measurement and Control Datalogger
RetiredNotes—
CR5000
Measurement and Control Datalogger
RetiredNotes—
CR9050
5-Volt Analog Input Module
RetiredNotes—
Datalogger for Measurement and Control
Datalogger for Measurement and Control
RetiredNotes—
Micrologger for Data Acquisition
Micrologger for Data Acquisition
RetiredNotes—
Micrologger®
Micrologger®
RetiredNotes—
Related Resources
Product Life Events
- Available for Order